79-year-old woman. Syncope
She has a permanent pacemaker implanted several years ago.
Stage 1 of 2
Her triage ECG is recorded.
Stage 2 of 2
The pacemaker clinic confirms an AAI device implanted eight years ago for sinus node dysfunction, with normal atrial capture and sensing and adequate battery life. She takes bisoprolol 2.5 mg daily. Her clinic ECG eighteen months ago showed 1:1 conduction with a narrow QRS.
Atrial pacing with Mobitz type I AV block, RBBB and right axis deviation
This tracing shows atrial pacing with Mobitz type I second-degree AV block. Pacing spikes precede every P wave, confirming normal atrial capture, but no ventricular spikes are present because this is a single-chamber atrial device. The PR lengthens progressively until a paced P fails to conduct, producing grouped beating at a ventricular rate of 53.
The QRS shows right bundle branch block with marked right axis deviation, consistent with left posterior fascicular block and so with probable bifascicular disease. Development of AV block is the recognised long-term limitation of single-chamber atrial pacing for sinus node dysfunction; in the DANPACE trial, atrial-only pacing carried twice the risk of pacemaker reoperation compared with dual-chamber pacing. The management is upgrade to a dual-chamber pacemaker. Rate-limiting drugs should be reviewed, but the structural conduction disease requires a ventricular lead.