66-year-old woman. Monitor alarm during a laparoscopic operation
Dual-chamber pacemaker implanted four years ago for complete heart block after aortic valve replacement, pacing-dependent at her last device check. Bisoprolol and apixaban, the apixaban held for surgery.
On arrival
HR
42
SpO₂ (%)
96
Respiratory rate
12
Temperature (°C)
36.1
Systolic BP (mmHg)
78
Diastolic BP (mmHg)
46
Stage 1 of 2
Eleni Vasilakis is forty minutes into a laparoscopic cholecystectomy under general anaesthesia. As the surgeon works on the gallbladder bed with monopolar diathermy, the monitor alarms and the pulse oximeter trace flattens for a few seconds at a time before recovering. She is ventilated, the end-tidal carbon dioxide dips with each episode, and the surgeon has not noticed. This is the rhythm strip recorded across one of them.
The diathermy is stopped and a magnet is taped over the generator. The oximetry trace steadies and the end-tidal carbon dioxide returns to its previous level. The surgeon resumes with short bursts while the strip runs.
Pacemaker oversensing of diathermy causing ventricular asystole
Electromagnetic interference from monopolar diathermy is the commonest intraoperative cause of pacemaker oversensing. The device interprets the high-frequency signal as cardiac electrical activity and withholds its output, producing ventricular asystole in a pacing-dependent patient. The distinction from failure to capture rests on one observation: whether pacing spikes appear during the pauses. No spikes means the device chose not to fire.
A magnet switches the device to asynchronous mode, restoring pacing regardless of what it senses. Bipolar diathermy produces less electromagnetic spread and is preferred near an implanted device. The device should be interrogated after surgery to confirm that sensing and thresholds are unchanged.
References
Crossley GH, Poole JE, Rozner MA, et al. The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors. Heart Rhythm. 2011;8(7):1114-1154.
<iframe src="https://www.squiggler.io/blog/monitor-alarm-during-a-laparoscopic-operation/embed" title="Monitor alarm during a laparoscopic operation — ECG case from Squiggler" width="100%" height="900" style="border:0" loading="lazy"></iframe>
This tracing was generated, not recorded. It opens in the studio with
every dial live.
Comments are read before they appear. Disagreement about a reading is welcome —
this is a teaching column and the arguments are the useful part. No account, no email.