Monitor alarm during a laparoscopic operation

The case

66-year-old woman. Monitor alarm during a laparoscopic operation

Dual-chamber pacemaker implanted four years ago for complete heart block after aortic valve replacement, pacing-dependent at her last device check. Bisoprolol and apixaban, the apixaban held for surgery.

On arrival

HR
42
SpO₂ (%)
96
Respiratory rate
12
Temperature (°C)
36.1
Systolic BP (mmHg)
78
Diastolic BP (mmHg)
46

Stage 1 of 2

Eleni Vasilakis is forty minutes into a laparoscopic cholecystectomy under general anaesthesia. As the surgeon works on the gallbladder bed with monopolar diathermy, the monitor alarms and the pulse oximeter trace flattens for a few seconds at a time before recovering. She is ventilated, the end-tidal carbon dioxide dips with each episode, and the surgeon has not noticed. This is the rhythm strip recorded across one of them.

Tracing 1 of 2 for the quiz “Monitor alarm during a laparoscopic operation”.
Monitor strip recorded during the procedure, lead II, 25 mm/s, 10 mm/mV. Open this tracing in the studio
What does the strip show?

The strip shows paced complexes interrupted by prolonged pauses during which no pacing spikes appear. In a pacing-dependent patient with complete heart block, the absence of both spikes and QRS complexes during these intervals means the device is withholding its output. A fractured lead would produce continuous loss of capture or sensing, not intermittent pauses coinciding with diathermy use.

Which feature on the paper separates these two mechanisms?

The critical distinction is the presence or absence of pacing spikes during the pauses. In failure to capture, the device fires on schedule but the myocardium does not respond, producing visible spikes with no following QRS. Here the pauses contain no spikes at all, confirming that the generator has withheld its output entirely. It has interpreted the diathermy signal as native cardiac activity and inhibited itself.

She has had three of these episodes in five minutes. What do you do?

A magnet over a pacemaker generator switches it to asynchronous mode, in which it paces at a fixed rate regardless of what it senses, eliminating oversensing as a cause of inhibition. Atropine is ineffective because the problem is not vagal tone but an absent pacing output in a patient with complete heart block and no native escape. Continuing without intervention risks prolonged asystole.

Stage 2 of 2

The diathermy is stopped and a magnet is taped over the generator. The oximetry trace steadies and the end-tidal carbon dioxide returns to its previous level. The surgeon resumes with short bursts while the strip runs.

Tracing 2 of 2 for the quiz “Monitor alarm during a laparoscopic operation”.
Monitor strip after the magnet was applied, lead II, 25 mm/s, 10 mm/mV. Open this tracing in the studio

Observations

HR
85
SpO₂ (%)
99
Respiratory rate
12
Temperature (°C)
36.1
Systolic BP (mmHg)
106
Diastolic BP (mmHg)
62
What does the second strip confirm?

The second strip shows regular paced complexes at a fixed rate of about 85 with no pauses, even though diathermy has resumed. This confirms asynchronous pacing: the device is firing on its own clock and ignoring all sensed input. The QRS morphology is unchanged from the first strip, confirming the same ventricular activation from the same lead, and the rate is the magnet rate rather than a reprogrammed setting.

Answer every question to see the diagnosis.

Diagnosis

Pacemaker oversensing of diathermy causing ventricular asystole

Electromagnetic interference from monopolar diathermy is the commonest intraoperative cause of pacemaker oversensing. The device interprets the high-frequency signal as cardiac electrical activity and withholds its output, producing ventricular asystole in a pacing-dependent patient. The distinction from failure to capture rests on one observation: whether pacing spikes appear during the pauses. No spikes means the device chose not to fire.

A magnet switches the device to asynchronous mode, restoring pacing regardless of what it senses. Bipolar diathermy produces less electromagnetic spread and is preferred near an implanted device. The device should be interrogated after surgery to confirm that sensing and thresholds are unchanged.

References

  1. Crossley GH, Poole JE, Rozner MA, et al. The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors. Heart Rhythm. 2011;8(7):1114-1154.
Embed
<script async src="https://www.squiggler.io/embed.js" data-case="monitor-alarm-during-a-laparoscopic-operation"></script>

Or, where scripts are stripped:

<iframe src="https://www.squiggler.io/blog/monitor-alarm-during-a-laparoscopic-operation/embed" title="Monitor alarm during a laparoscopic operation — ECG case from Squiggler" width="100%" height="900" style="border:0" loading="lazy"></iframe>

This tracing was generated, not recorded. It opens in the studio with every dial live.

Open the studio

Discussion

Comments are read before they appear. Disagreement about a reading is welcome — this is a teaching column and the arguments are the useful part. No account, no email.